The HiTemp140 is a submersible, high temperature data logger built for use in harsh environments. This best-seller is extremely versatile and can be tailored to the application with various probe lengths and styles available.

The HiTemp140 can withstand temperatures ranging from -40 °C to +140 °C (-40 °F to +284 °F), but is capable of measuring extended temperatures up to 260 °C (500 °F) (probe dependent). With a storage capacity of up to 65,536 time and date stamped readings, the HiTemp140 can be programmed for immediate start or delayed start up to 18 months in advance.

The HiTemp140 is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

Body of data logger cannot exceed 140 °C (284 °F), but can extend to higher temperatures when used with MadgeTech’s thermal barrier options.

HiTemp140-FR fast response probe is very thin to increase the response time, hence fragile, should be used carefully to avoid physical damages, which is not covered under manufacturer warranty.

General

Memory65,536 readings
Reading Rate1 reading every second up to 1 reading every 24 hours
Baud Rate125,000
Battery Life1 year typical (1 minute reading rate at 25 °C/77 °F)
Fast reading intervals and extreme temperatures will shorten the battery life substantially
User Replaceable Battery
Material316 Stainless Steel
HiTemp 140 Rigid and Rigid Transitional Diameter DimensionsHiTemp140-1 Body: 1.55 in x 0.97 in dia. (40 mm x 25 mm dia.)
Probe: 1 in x 0.125 in dia. (0.188 in transitional dia.)
25 mm x 3.2 mm dia. (4.8 mm transitional dia.)
Body (for 2 in, 5.25 in, 7 in;): 1.75 in x 0.97 in dia. (45 mm x 25 mm dia.)HiTemp140-2 : 2.0 in x 0.188 in dia. (51 mm x 4.8 mm)HiTemp140-2-TD: 2.0 in x 0.125 in dia. (0.188 in transitional dia.)
(51 mm x 3.2 mm dia. (4.8 mm transitional dia.)HiTemp140-5.25: 5.25 in x 0.188 in dia. (134 mm x 4.8 mm)HiTemp140-5.25-TD: 5.25 in x 0.125 in dia. (0.188 in transitional dia.)
(134 mm x 3.2 mm dia. (4.8 mm transitional dia.)HiTemp140-7: 7.0 in x 0.188 in dia. (178 mm x 4.8 mm dia.)
HiTemp140 Flexible Probe DimensionsBody: 2.95 in x 0.97 in x 0.97 in (75 mm x 24.6 mm x 24.6 mm)
Probes: HiTemp140-FP-6: 6 in x 0.10 in (152.5 mm x 2.5 mm)
HiTemp140-FP-12: 12 in x 0.10 in (304.8 mm x 2.5 mm)
HiTemp140-FP-36: 36 in x 0.10 in (914 mm x 2.5 mm)
HiTemp140-FP-72: 72 in x 0.10 in (1829 mm x 2.5 mm
HiTemp140 Pig Tail Probe DimensionsBody: 1.75 in x 0.97 in dia. (45 mm x 25 mm dia.)

HiTemp140-PT-1: Probe tip is 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.)
Bendable portion is 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)

HiTemp140-PT-5: Probe tip is 4.8 in x 0.125 in dia. with 1 in x 0.188 dia. handle (121 mm x 3.2 mm dia. with 25 mm x 4.8 mm dia. handle)
Bendable portion is 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)

HiTemp140 Fast Response Probe DimensionsBody: 1.89 in x 0.97 in x 0.97 in (48.1 mm x 24.6 mm x 24.6 mm)
Probe: 1.75 in x 0.0625 in dia. (0.188 in transitional dia.) (44.5 mm x 1.59 mm (4.78 mm Transitional dia.))
IP RatingIP68
SubmersibleYes
Operating Environment-40 °C to +140 °C (-40 °F to +284 °F)
0 %RH to 100 %RH (Body of Logger Cannot Exceed 140 °C)
Required Interface PackageIFC400
ApprovalsCE

 

Temperature

Temperature Sensor100 Ω Platinum RTD
Temperature Resolution0.01 °C (0.018 °F)
Temperature Calibrated Accuracy±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F)
±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F)
±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to +-4.02 °F)
Probe Temperature Range-200 °C to +260 °C (-328 °F to +500 °F) (Body of Logger Cannot Exceed 140 °C)

HiTemp140 Fast Response Probe Specification Sheet

HiTemp140 Fast Response Probe Product User Guide

HiTemp140 Fast Response Probe Dimensional Drawing

 

HiTemp140 Response Time Chart

Probe TypeMaterialOperating EnvironmentProbe Measurement Range
RigidMetalTemperature: -40 °C to +140 °C (-40 °F to +284 °F)
Humidity: 0 %RH to 100 %RH
Pressure: 0.002 PSIA to 100 PSIA
Temperature: -200 °C to +260 °C (-328 °F to +500 °F)
Probe only. Body must not exceed Operating Environment
Rigid Transitional Diameter (TD)MetalTemperature: -40 °C to +140 °C (-40 °F to +284 °F)
Humidity: 0 %RH to 100 %RH
Pressure: 0.002 PSIA to 100 PSIA
Temperature: -200 °C to +260 °C (-328 °F to +500 °F)
Probe only. Body must not exceed Operating Environment
Flexible (FP)Bendable PTFETemperature: -40 °C to +140 °C (-40 °F to +284 °F)
Humidity: 0 %RH to 100 %RH
Pressure: 0.002 PSIA to 60 PSIA
Temperature: -60 °C to +260 °C (-76 °F to 500 °F)
Probe only. Body must not exceed Operating Environment
PTBendable MetalTemperature: -40 °C to +140 °C (-40 °F to +284 °F)
Humidity: 0 %RH to 100 %RH
Pressure: 0.002 PSIA to 100 PSIA
Temperature: -200 °C to +350 °C (-328 °F to +662 °F)
Probe only. Body must not exceed Operating Environment
Fast Response (FR)MetalTemperature: -40 °C to +140 °C (-40 °F to +284 °F)
Humidity: 0 %RH to 100 %RH
Pressure: 0.002 PSIA to 100 PSIA
Temperature: -200 °C to +260 °C (-328 °F to +500 °F)
Probe only. Body must not exceed Operating Environment